MAINTENANCE
Milli-Q Synthesis/Synthesis A10 62
Milli-Q Synthesis A10 only: When you ‘Close The Valve’, the Milli-Q will go to PRE OPERATE Mode
and then will start an automatic 60 minute A10 CLEANING.
The A10 CLEANING is a way of flushing
out any particles from the A10 that may have been present.
Milli-Q Synthesis A10 only: Allow the A10 CLEANING to count down to 0 minutes.
The remaining steps of this section apply to both the Milli-Q Synthesis and the Milli-Q Synthesis A10.
Any remaining air in Quantum Cartridge can be ‘hand purged’ out now.
Get a screwdriver with a very thin shaft and small head.
Locate the small hole on the top part of the Quantum Cartridge Door.
Gently insert the screwdriver into the Quantum Cartridge. This will open a small vent that will expel any air in
the Quantum Cartridge.
Be very careful when pushing the tip of the screwdriver into the
Vent Hole of the Quantum Cartridge. The Vent Hole only needs to
be opened a very small amount.
If all of the air does not expel out, do not worry. You can repeat this step later after more water has flowed
through the Milli-Q. Remember to have the Milli-Q in STANDBY or PRE OPERATE Mode.
You now need to allow the Q-Gard Pack and the Quantum Cartridge to hydrate
overnight. This helps to produce high quality Product Water.
At this time, the Millipak is not installed. The Barbed Fitting and Tygon Tubing should still be attached to the
end of the POU Dispenser.
Leave the Milli-Q in PRE OPERATE Mode overnight or for several hours. This will aid in rinsing off
the purification media inside the Q-Gard Pack and the Quantum Cartridge.
(The next morning) Bring the POU Dispenser Trigger forward to put the Milli-Q into PRODUCT Mode.
Allow about 10 to 15 Litres (7 to 10 minutes of PRODUCT Mode) of water to be dispensed from the Milli-Q.
Bring the POU Dispenser Trigger upright. The Milli-Q will go into PRE OPERATE Mode after 90 seconds.
Leave the Milli-Q in PRE OPERATE Mode.
The Q-Gard Pack and Quantum Cartridge are now fully air purged and hydrated.
Proceed to Section 6-4 for instructions on installing the Millipak.
P R E O P E R A T E
A 1 0 C L E A N I N G : 6 0
S T A ND B Y
C L O S E T H E V A L V E
ATTENTION