P214 4 mA TRIM
Use this feature (in combination with 20 mA Trim) if the device connected to the mA output
displayed is out of calibration, yet device recalibration is impractical.
Adjust this value (preset to 0.000) as required to make the device connected
indicate 4.000 mA when this parameter is accessed.
values: -1.00 to 1.000
P215 20 mA TRIM
Use this feature (in combination with 4 mA Trim) if the device connected to the mA output
displayed is out of calibration, yet device recalibration is impractical.
Adjust this value (preset to 0.000) as required to make the device connected
indicate 20.00 mA when this parameter is accessed.
values: -1.00 to 1.000
P219 mA FAILSAFE
Use this feature if mA output failsafe operation, independent from the Failsafe Material
Level (P071) is desired.
To select an independent mA Failsafe option value:
1. , to display the Auxiliary Function symbol,
2. or , to scroll access the failsafe options.
3. , with the desired option displayed.
values: "OFF" to have the mA output respond to the Failsafe Material Level (P071).
"HOLd" to hold the output at the "last known" value until normal operation resumes.
"LO" to produce the "Empty" mA output immediately on
Failsafe Timer (P070) expiry.
"HI" to produce the "Span" mA output immediately on
Failsafe Timer (P070) expiry.
Alternatively, to produce an immediate mA output at a specific value, enter the value required.
values: 0.000 to 22.00
%
PL-508 44