PROFILE RECORDS (P330 to P337)
The following parameters are for authorized Milltronics service personnel or Instrumentation Technicians
familiar with Milltronics echo processing techniques.
Use these features to record and save a total of up to 10 Echo profiles, initiated manually (P330), or
automatically (P331 et al). See Scope displays (P810) for echo profile viewing hardware / software
requirements.
If 10 Echo Profiles are already saved, addresses 1 through 10 are filled, the oldest automatically initiated
record is overwritten. Manually initiated records are not automatically overwritten. All records are
automatically deleted in the event of a power interruption.
When a record is displayed, results are based on current programming (which may have been altered
since the record was saved). This permits the effect on the echo profile to be observed when changing
an echo parameter.
P330 PROFILE RECORD
In addition to serving as a library for profile records, this parameter provides two functions:
» manually records and saves echo profiles
» displays an echo profile, recorded manually or automatically, e.g. via an oscilloscope.
To select a record address:
until address index is enabled.
selects the desired address, 1 to 10, and displays the associated parameter
value: ‘- - - -’ = no record
‘ x# ’ = record
where: x = A, automatically initiated
= U, manually initiated
# = transducer number
To manually record a profile:
transducer* fires and echo profile is
recorded into the internal scope buffer for display.
e.g. initial parameter display
e.g. address 2 selected, no record saved