P840 SHORT SHOT NUMBER
Enter the number of short shots to be fired (and results averaged) per transmit pulse.
values: 0 to 100
P841 LONG SHOT NUMBER
Enter the number of long shots to be fired (and results averaged) per transmit pulse.
This value is automatically altered by Measurement Response (P003).
(See Measurement Response).
values: 0 to 200
P842 SHORT SHOT FREQUENCY
Use this feature to adjust the short shot transmit pulse frequency (in kHz).
This feature is automatically altered when Transducer (P004) is altered.
values: 10.00 to 60.00
P843 LONG SHOT FREQUENCY
Use this feature to adjust the long shot transmit pulse frequency (in kHz).
This feature is automatically altered when Transducer (P004) is altered.
values: 10.00 to 60.00
P844 SHORT SHOT WIDTH
Use this feature to adjust the width (in ms) of the short shot transmit pulse.
This feature is automatically adjusted by Transducer (P004).
values: 0.000 to 5.000
P845 LONG SHOT WIDTH
Use this feature to adjust the width (in ms) of the long shot transmit pulse.
This feature is automatically adjusted by Transducer (P004).
values: 0.000 to 5.000
TECHNICAL REFERENCE
PL-508 70