Setup Menu System Interface
4 - 42 Operator’s Manual of Anesthesia System
MENU OPTION DESCRIPTION
Ethernet
MAC Address
Displays the MAC address of the anesthesia
system.
Device Name Sets the device name.
Obtain IP Address
Automatically
Enables or disables the Obtain IP Address
Automatically feature. When the feature is
enabled, the [IP Address], [Subnet] and [Default
Gateway] cannot be manually changed. When
the feature is disabled, the [IP Address], [Subnet]
and [Default Gateway] need to be manually
changed. In this case, the Obtain DNS Server
Address Automatically and Obtain Default DNS
Suffix Automatically features, and the Obtain IP
Address Automatically feature in the SNTP menu
are disabled.
IP Address Sets the IP address.
Subnet Mask Sets the subnet mask.
Default Gateway Sets the default gateway.
Obtain DNS Server
Address
Automatically
Enables or disables the Obtain DNS Server
Address Automatically feature. When the feature
is enabled, the [Preferred DNS Server] and
[Alternate DNS Server] cannot be manually
changed. When the feature is disabled, the
[Preferred DNS Server] and [Alternate DNS
Server] need to be manually changed.
Preferred DNS
Server
Sets the preferred DNS server.
Alternate DNS
Server
Sets the alternate DNS server.
Serial
Protocol
Sets the serial communication protocol.
NOTE: If the protocol is set to MR-
WATO, the anesthesia system
supports communications
with a Mindray patient
monitor with the Mindray
Benelink module.
Baud Rate Sets the Baud rate.
Data Bits Sets the data bits.
Stop Bits Sets the stop bits.
Parity Sets the parity.
Data Interval Sets the data intervals.
HL7 Protocol
Version
Sets the HL7 protocol version.
Compatibility Sets the serial communication protocol version.
Send Alarms Enables or disables the Send Alarms feature.
Send Alarm Ack Enables or disables the Send Alarm Ack feature.
Table 4-9 Network menu