Operator’s Manual of Anesthesia System 8 - 15
Alarms and Messages Alarms and Prompt Messages
EtN
2
O
Too High
EtN
2
O > high
alarm limit.
1. Check alarm limit:
EtN
2
O High.
2. Check fresh gas
balance gas setting.
Medium No No Yes No No
EtN
2
O
Too Low
EtN
2
O < low
alarm limit.
1. Check alarm limit:
EtN
2
O Low.
2. Check whether the
sampleline is
disconnected.
3. Check fresh gas N
2
O
setting, and the N
2
O
source is connected
correctly.
Medium No Yes Yes No No
FiN
2
O Too
High
FiN
2
O > high
alarm limit.
1. Check alarm limit:
FiN
2
O High.
2. Check fresh gas
balance gas setting.
Medium No No Yes No No
FiN
2
O Too
Low
FiN
2
O < low
alarm limit.
1. Check alarm limit:
FiN
2
O Low.
2. Check whether the
sampleline is
disconnected.
3. Check fresh gas N
2
O
setting, and the N
2
O
source is connected
correctly.
Medium No Yes Yes No No
EtHAL
Too High
EtHAL > high
alarm limit
setting.
1. Check alarm limit:
EtHal High.
2. Check AA% setting.
Medium No No Yes No No
EtHAL
Too Low
EtHAL < low
alarm limit
setting.
1. Check alarm limit:
EtHal Low.
2. Check AA% setting.
3. Check whether the
sampleline is
disconnected.
Medium No Yes Yes No No
FiHAL Too
High
FiHAL > high
alarm limit
setting.
1. Check alarm limit:
FiHal High.
2. Check AA% setting.
Medium No No Yes No No
FiHAL Too
Low
FiHAL < low
alarm limit
setting.
1. Check alarm limit:
FiHal Low.
2. Check AA% setting.
3. Check whether the
sampleline is
disconnected.
Medium No Yes Yes No No
EtENF Too
High
EtENF > high
alarm limit
setting.
1. Check alarm limit:
EtEnf High.
2. Check AA% setting.
Medium No No Yes No No
INFORMA-
TION CAUSE ACTION
ALARM
PRIORITY
DISABLE
WHEN
ALARM
IS OFF
DISABLE
IN
CARDIAC
BYPASS
MODE
DISABLE
IN
STANDB
Y MODE
DISABLE
IN
MONITO
R MODE
DISABLE
IN ACGO
MODE
Table 8-3 Physiological Alarm Messages

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
BrandMindray
ModelA9
CategoryMedical Equipment
LanguageEnglish

Related product manuals