Alarms and Prompt Messages Alarms and Messages
8 - 16 Operator’s Manual of Anesthesia System
EtENF Too
Low
EtENF < low
alarm limit
setting.
1. Check alarm limit:
EtEnf Low.
2. Check AA% setting.
3. Check whether the
sampleline is
disconnected.
Medium No Yes Yes No No
FiENF Too
High
FiENF > high
alarm limit
setting.
1. Check alarm limit:
FiEnf High.
2. Check AA% setting.
Medium No No Yes No No
FiENF Too
Low
FiENF < low
alarm limit
setting.
1. Check alarm limit:
FiEnf Low.
2. Check AA% setting.
3. Check whether the
sampleline is
disconnected.
Medium No Yes Yes No No
EtISO Too
High
EtISO > high
alarm limit
setting.
1. Check alarm limit:
EtIso High.
2. Check AA% setting.
Medium No No Yes No No
EtISO Too
Low
EtISO < low
alarm limit
setting.
1. Check alarm limit:
EtIso Low.
2. Check AA% setting.
3. Check whether the
sampleline is
disconnected.
Medium No Yes Yes No No
FiISO Too
High
FiISO > high
alarm limit
setting.
1. Check alarm limit:
FiISO High.
2. Check AA% setting.
Medium No No Yes No No
FiISO Too
Low
FiISO < low
alarm limit
setting.
1. Check alarm limit:
FiIso Low.
2. Check AA% setting.
3. Check whether the
sampleline is
disconnected.
Medium No Yes Yes No No
EtSEV Too
High
EtSEV > high
alarm limit
setting.
1. Check alarm limit:
EtSev High.
2. Check AA% setting.
Medium No No Yes No No
EtSEV Too
Low
EtSEV < low
alarm limit
setting.
1. Check alarm limit:
EtSev Low.
2. Check AA% setting.
3. Check whether the
sampleline is
disconnected.
Medium No Yes Yes No No
FiSEV Too
High
FiSEV > high
alarm limit
setting.
1. Check alarm limit:
FiSev High.
2. Check AA% setting.
Medium No No Yes No No
FiSEV Too
Low
FiSEV < low
alarm limit
setting.
1. Check alarm limit:
FiSev Low.
2. Check AA% setting.
3. Check whether the
sampleline is
disconnected.
Medium No Yes Yes No No
INFORMA-
TION CAUSE ACTION
ALARM
PRIORITY
DISABLE
WHEN
ALARM
IS OFF
DISABLE
IN
CARDIAC
BYPASS
MODE
DISABLE
IN
STANDB
Y MODE
DISABLE
IN
MONITO
R MODE
DISABLE
IN ACGO
MODE
Table 8-3 Physiological Alarm Messages

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
BrandMindray
ModelA9
CategoryMedical Equipment
LanguageEnglish

Related product manuals