Operator’s Manual of Anesthesia System 8 - 19
Alarms and Messages Alarms and Prompt Messages
BFCS self-
test error
BFCS self-test error 1. Restart machine
to repeat the test.
2. Please contact
Mindray Technical
Support if the
problem persists.
High Startup All
Aux Control
Module
Selftest Error
1. Board self-test error.
2. After being powered
on, the CPU board
failed to communicate
with the auxiliary
monitoring board.
1. Repeat the test.
2. Please contact
Mindray Technical
Support if the
problem persists.
High Startup Non-
Functional
Auxi Ctrl
Module Self
Test: Time out
Self-test result
cannot be obtained
due to an internal
communication error.
High Startup Non-
Functional
Ventilator
Selftest Error
1. Board self-test error.
2. After being powered
on, the CPU board
failed to communicate
with the ventilator.
1. Repeat the test.
2. Please contact
Mindray Technical
Support if the
problem persists.
High Startup Non-
Functional
Ventilator
Control Board
Self Test:
Time out
Self-test result
cannot be obtained
due to an internal
communication error.
High Startup Non-
Functional
Ventilator
Voltage Error
Ventilator voltage
error.
High Startup Manual
Only
PEEP Valve
Failure
1. PEEP valve voltage
error.
2. PEEP valve pressure
error.
1. Repeat the test.
2. Please contact
Mindray Technical
Support if the
problem persists.
Medium Startup Manual
Only
Insp Valve
Failure
1. Inspiratory valve
voltage error.
2. Inspiratory valve
flow error.
1. Repeat the test.
2. Please contact
Mindray Technical
Support if the
problem persists.
Medium Startup Manual
Only
Safety Valve
Failure
Safety valve voltage
error.
1. Repeat the test.
2. Please contact
Mindray Technical
Support if the
problem persists.
Medium Startup Manual
Only
Auto/Manual
Switching
Failure
Auto/Manual position
switch failure
1. Repeat the test.
2. Please contact
Mindray Technical
Support if the
problem persists.
High Startup Manual
Only
Auto/manual valve
fault
High Startup Non-
Functional
Flow Sensor
Failure
Ventilator flow is out of
range.
1. Repeat the test.
2. Please contact
Mindray Technical
Support if the
problem persists.
Low Startup Manual
Only
INFORMA-
TION CAUSE ACTION
ALARM
PRIORITY
MACHINE
MODE
WHEN
CHECKED
STARTUP
RESULT IF
FAIL
Table 8-4 Startup Alarm Messages

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA