Operator’s Manual of Anesthesia System 8 - 29
Alarms and Messages Alarms and Prompt Messages
8.8.2.7 AGSS Alarms
8.8.2.8 HFNC Module Alarms
8.8.3 Prompt Message
8.8.3.1 Prompt Messages Displayed in Alarms Area
INFORMA-
TION CAUSE ACTION
ALARM
PRIORITY
MACHINE
MODE
WHEN
CHECKED
DISABLED
IN
STANDBY
MODE
Scavenging
Flow is Too
High
The AGSS flow rate is
higher than the upper
limit.
Set the vacuum
flow so that the
float position is
between the Min
and Max lines.
Low Runtime No
AGSS Failure AGSS failure. Please contact
Mindray Technical
Support.
Medium Runtime No
Table 8-10 AGSS Alarms
INFORMA-
TION CAUSE ACTION
ALARM
PRIORITY
MACHINE
MODE
WHEN
CHECKED
DISABLED
IN
STANDBY
MODE
HFNC
Module
Failure
1. O
2
branch flow
sensor error.
2. Air branch flow
sensor error.
Please contact
Mindray Technical
Support.
Medium Runtime No
HFNC
Selftest Error
1. EEPROM self-test
error.
2. O
2
proportional
valve self-test error.
3. Air proportional
valve self-test error.
4. Zero reading error.
1. Restart HFNC.
2. Please contact
Mindray Technical
Support if the
problem persists.
Medium Runtime No
Calibrate
HFNC
Module
1. Calibration table
isn't found in EEPROM.
2. Checksum of the
calibration table does
not match.
Please contact
Mindray Technical
Support.
Low Startup No
Table 8-11 HFNC Module Alarms
INFORMATION REMARK
Volume and Apnea
Alarms are OFF
This message appears when the [Alarms] is set to [Off] in the
Manual mode.
CO
2
and CO
2
Apnea
Alarms are OFF
This message appears when [CO
2
Alarms] is set to [Off] in the
Manual mode.
Load Configuration
Failure
This message appears when loading user or latest configuration failed.
Save Profile Failure This message appears when an error occurs during saving user profiles.
Table 8-12 Prompt Messages Displayed in Alarms Area

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA

Related product manuals