Description of Self-test Test Items  C-1 
Appendix C  Description of 
Self-test Test 
Items 
C.1.1  Z0101 Hard Disk Verify Test 
  Top test items 
N/A 
  Test content 
Traverse all hard disk files from Doppler installation directory, compare and verify the 
files with the archived hard disk data. 
  Analysis to test failure 
If the system prompts “Failed to open the result file”, the verification file does not exist. 
If the system prompts “The failed CRC: current verification value (failed file path, 
correct verification value)”, the hard disk data is damaged. 
  Suggestion to test failure 
Restore the hard disk data; replace the file from M9 directory in C local disk with the 
CRC_Result.txt from the restore package. 
C.1.2  Z0201 PC Module and DSP FPGA Interconnection 
Test 
  Top test items 
N/A 
  Test content 
Test whether PCIe communication between PC module and DSP FPGA works well. 
  Analysis to test failure 
Check whether Windows device manager recognizes Ultrasound Backbone Device. If 
the drive is not installed properly, it cannot be recognized. 
The drive goes wrong if the test result appears Error. 
PC module and DSP FPGA has communication error if the test result is FAIL. 
  Suggestion to test failure 
a)  Restore the device if the drive is not recognized. 
b)  Restart the device and perform the self-test if the test result appears Error. It is 
necessary to restore the device if Error re-appears. 
c)  It is recommended to replace CPU if the test result is FAIL. 
C.1.3  Z0202 DSP FPGA DDRIII Test 
  Top test items 
PC Module and DSP FPGA Interconnection Test 
  Test content