Product Specifications 2-13
WARNING:
1.
Be sure to connect the equipotential wire before inserting the
power plug into the receptacle; be
sure to pull out the power
plug from the receptacle before disconnecting the equipotential
wire; otherwise electric shock may result.
2.
When you connect another device to this system, you should
use the equipotential wire to connect each of equipotential
terminals; otherwise electric shock may result.
3. Connect the e
arth cable before turning ON the system.
Disconnect the earth cable after turning OFF the system.
Otherwise electric shock may result.
4.
DO NOT connect this system to outlets with the same circuit
breakers and fuses that control the current to devices such as
life-support systems. If this system malfunctions and generates
over current, or when there is an instantaneous current at power
ON, the circuit breakers and fuses of the building’s supply
circuit may be tripped.
2.1.4 Peripherals Supported
The peripheral devices supported by the system. The information is shown as below:
No. Item Model
1.
Graph/text printer
HP Officejet 7000 wide format
2. Black/white video printer
MITSUBISHI P95DW-N
SONY UP-D897
SONY UP-D898MD
SONY UP-X898MD
3. Digital color video printer
SONY UP-D25MD
4. 1D/2DBarcode reader
SYMBOL DS4308
SYMBOL LS2208
JADAK HS-1M
JADAK HS-1R
5. DVR
6. External DVD R/W drive ASUS DVD SDRW-08D2S-U
NOTE: printer UP-D898MD and UP-X898MD should be used in compatible with 1.00.7 2015-1-27
OS version or above versions. You can also select “DVR: 897” in the DIGITAL -> - DRIVER path on
the printer (use UPD897 driver) if OS is not updated.
WARNING:
This device complies with IEC60601-1-2:2014, and its RF emission
meets the requirements of CISPR11 Class B.
In a domestic
environment, the customer or the user should guarantee to connect
the system with Class B peripheral devices; otherwise RF
interference may result and the customer or the user must take
adequate measures accordingly.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray M9T and is the answer not in the manual?

Mindray M9T Specifications

General IconGeneral
BrandMindray
ModelM9T
CategoryMedical Equipment
LanguageEnglish

Related product manuals