3-2 System Installation
3.1.2 Installation Condition
3.1.2.1 Space Requirements
Place the system with the necessary accessories at a proper position for convenient use.
Place the system in a room with good ventilation or having an air conditioning unit.
Maintain a generous – free air flowing space around the back and both sides of the system; failure
may result due to increased rise in system operating temperature.
A combination lighting system in the room (dim/bright) is recommended.
Except the receptacle dedicated for the ultrasound system, at least 3-4 spare receptacles on the
wall are available for the other medical devices and peripheral devices.
Power outlet and place for any external peripheral are within 2 m of each other with peripheral
within 1 m of the unit to connect cables.
3.1.2.2 Network Environment
Both wireless and wired LAN functions are supported by this ultrasound device.
Data transmission is allowed between different departments or areas without network cable.
Network can be automatically connected after disconnection in case that the device is required to
be moved, wireless transmission task can be recovered after the network resumed to normal
condition. Confirm the network devices and network conditions before the installation.
General information: default gateway IP address, and the other routers related information.
DICOM application information: DICOM server name, DICOM port, channels, and IP address.
3.1.3 Installation Confirmation
Please confirm the following items before installation:
The video format of installation area or country.
The language of installation area or country.
Power frequency of installation area or country.
The universal obstetrics formula and other measurement formula of installation area or country.
The preset values of installation area or country that are different from the default values.
The doctor’s operation habits.
The items above prior to the installation training, and do the system settings according to the
universal setting of installed region or country.
3.2 Unpacking
Unpacking tool: scissor
Installation duration: 2 people, 10 minutes.
3.2.1 Unpacking Process
3.2.1.1 Trolley Unpacking
1. Use the scissor to cut off four white tapes around the external package as follows: