EasyManua.ls Logo

Minebea Intec CAS-SPC Series - Data Output Formats; Data Output Format Overview

Minebea Intec CAS-SPC Series
188 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Data Output Format
You can output the value displayed in the measured value line and the weight unit,
with or without a data ID code. Whether the data ID code is included in the output
depends on your settings under “Line Format.“
Examples + 235 pcs without ID code
Qnt + 235 pcs with ID code
“Line Format“ settings: For raw data (16 characters): without header,
for other apps. (22 characters): with “header“ (factory setting).
Data Output Format with 16 Characters
Display segments that are not activated are output as spaces.
Values with no decimal point are output without a decimal point.
The type of character that can be output depends on the character’s position:
Normal Operation
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
+ * A A A A A A A A * E E E CR LF
or * A A A A A A A A * E E E CR LF
or * * * * * * * * * * * * * CR LF
+-: Signs; the “+“ sign can be hidden under “
Device parameters:
COMx:Data communications:SBI:Sign format
.“
*: Space
A: Digit or letter (max. 7 characters plus decimal point)
E: Unit symbol
1)
(1 to 3 letters followed by 2-0 spaces)
CR: Carriage return
LF: Line feed
Special Codes
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
* * * * * * * – * * * * * CR LF
or * * * * * * * H * * * * * * CR LF
or * * * * * * * H H * * * * * CR LF
or * * * * * * * L * * * * * * CR LF
or * * * * * * * L L * * * * * CR LF
or * * * * * * * C * * * * * * CR LF
*: Space
– –: Final readout
H: Overload
HH: Overload in checkweighing
L: Underweight
L L: Underweight in checkweighing
C: Calibration/Adjustment
Error Messages
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
* * * E r r * * # # * * * * CR LF
* * * E r r * * # # # * * * CR LF
*: Space
#: Number (2 or 3 digit error number)
1
) Based on the model type, e.g. not all units are available for use in legal metrology
102 Operating Instructions Combics Complete Scales