Calibration, Adjustment
Purpose Perform calibration to determine the dierence between the value displayed and
the actual weight on the platform. Calibration does not entail making any changes
within the weighing equipment.
During adjustment, the dierence between the measured value displayed and the
true weight of a sample is corrected, or is reduced to an allowable level within
maximum permissible error limits.
Conguration for Use in Legal Metrology
Conguration of the weighing instrument for use in legal metrology is set by a
switch. The switch is located on the back of the weighing platform and covered by
a protective cap.
Using a veried scale in legal metrology in the EU:
The Type-Approval Certicate for veried scales is only valid for non-automatic
weighing instruments. For automatic operation with or without additional,
integrated equipment, please follow the applicable national regulations for the
installation location.
Externally Connected IS Scales: Before use in legal metrology, the scale should be adjusted via the internal
adjustment equipment at the installation location: see chapter “Conguring
Weighing Platforms,“ section “Internal Adjustment.“
3
The temperature range (°C) listed on the ID label should not be exceeded during
operation.
For Servicing: External adjustment for veried scales of accuracy class l
External adjustment is blocked in legal metrology (switch cover is sealed)
External adjustment is only possible by removing the seal. If the seal is broken,
the validity of verication will become void and you must have your scale
re-veried.
Using a veried scale in legal metrology with internal adjustment equipment:
t Before use in legal metrology, the “internal adjustment“ function should be
carried out at the installation location:
Opening the Menu Access Switch The menu access switch is located on the back of the indicator right next to the
weighing platform connection.
P
S
2
t Remove the cap.
t Slide the switch to the left (= “open“ position, for use in legal metrology).
Please refer Plates and Marking.
h
You can determine the position of the switch in the Setup menu under
Info:WP.“
86 Operating Instructions Combics Complete Scales