Delete preload
Setup Information – Deleting the preload when weighing in legal metrology is only possible when
the menu access switch is open.
– The J key must be assigned the “Delete preload“ function, see “Function
Allocation of the J Key for Linearization and Setting/Deleting the Preload.“
– Activating the display of geographical data has no eect on this function.
3
Once the preload has been deleted, the J key must be reallocated back to its
original function in the Setup menu, e.g. external calibration/adjustment with
default weights.
Procedure t For scales used in legal metrology, slide the menu access switch to the left
(= “open“ position).
t Press e turn the device back on.
y The Minebea Intec logo is displayed briey, after which the device is in normal
weighing mode.
t Remove the preload weight from the weighing platform.
y The display shows the removed preload weight as a negative value.
t Press J to start “Delete preload.“
y The display “CLR PREL“ appears briey.
After a short period of time the preload will be deleted and the indicator will
automatically switch back to weighing mode.
After the “Delete Preload” operation has been completed, the scale is zeroed.
When deleting the preload is complete, the GMP-compliant printout shown on the
left is generated.
t Slide the menu access switch to the right (= “closed“ position).
--------------------
24.02.2016 10:15
Model CAW3S
Ser.no. 12345678
Vers. C1 114.040216
BVers. 01-61-14
Model IS6DCS
Ser.no. C 12345678
--------------------
Delete preload
completed
--------------------
24.02.2016 10:20
Name:
--------------------
Operating Instructions Combics Complete Scales 51