Function Allocation of the J Key for Linearization and Setting/
Deleting the Preload
Purpose The J key is normally used for the calibration/adjustment function.
The following additional functions can be allocated to the key when the Service
mode is activated:
– External linearization with default weights
– External linearization with entered linearization weights
– Internal linearization (for external IS platforms only)
– Set preload (not possible in Legal Metrology)
– Delete preload (not possible in Legal Metrology)
3
Once linearization has been completed, or after a preload has been set or deleted
the function of the J key must be reallocated back to its original function in the
Setup menu, e.g. external calibration/adjustment with default weights.
Procedure t Remove the cap.
t Slide the menu access switch to the left (= “open“ position).
t Activate the Service mode, see “Service Mode.“
t Select weighing platform “
WP 1“ in the “Device Parameters“
menu item.
t If the “Internal“ setting is not already activated (marked by “o“), select
the setting using the “Q“ or “q“ soft key and conrm with “O.“
y The menu for the “WP-1 INTERNAL“ device parameters is displayed.
t Use the “Q“ or “q“ soft key to select and open the “Calibration/
adjustment
“ menu using “O.“
y The “CAL key function“ submenu is displayed.
t Use the “Q“ or “q“ soft key to select the corresponding menu item and conrm
with “l.“
y The menu item is marked by a circle “o.“
Note
The functions that can be congured in the “CAL key function“
submenu depends on the selected weighing platform and its conguration data.
Functions that cannot be accessed are not displayed in the selection screen.
Note
The function set in the “CAL key function“ menu is carried out in the
normal weighing mode because when you exit the Setup menu the Service mode is
deactivated. To perform the function on a digital weighing platform (such as an IS
platform), however, it must be carried out in Service mode.
t The procedure is as follows: after selecting the desired function and exiting
the Setup menu, reactivate Service mode again and then exit the Setup menu
immediately by pressing the M key or the “oo“ soft key.
y The scale is now in Service mode without this being displayed.
t Trigger the previously set function using the J key.
y The display shows “S_CAl:“ indicating that the scale is in Service mode.
t The corresponding menu item can be selected using the “Select“ soft key
and carried out using the J key.
t If you cancel the function using the ( key or by restarting the scale using
the e key, you will exit the Service mode.
46 Operating Instructions Combics Complete Scales