Application Initialization Data Which data is included in this block depends on the active application. In the
“Counting“ application, for example, the reference sample quantity and reference
weight are printed (plus a blank line).
nRef 10 pcs “Counting“ print block example:
wRef + 0.035 kg
Scale Identication
Ser.no. 1234567890 Print image example for weighing platform serial number:
Scale Identication This content is application-dependent. If provided in the application, the gross,
net and tare weights are usually printed, followed by a blank line. This block
is terminated by a dotted line.
G# + 1.402 kg “Counting“ application print image example:
T + 0.200 kg
N + 1.202 kg
Qnt 34 pcs
--------------------
GMP-compliant printouts
When this function, the printout is supplemented with a GMP header and a GMP
footer (GMP: “Good Manufacturing Practice”).
If the GMP-compliant printout is activated, the A symbol remains displayed until
the GMP footer is printed.
Setting: Setup menu under “Device parameters:Config.
printout:ISO/GLP/GMP- printout
.“
You can choose from the following settings:
– GMP-compliant printout o (factory setting)
– GMP-compliant printout for multiple application results
The GMP header is included from the rst printout generated subsequent to the
activation of the GMP printout.
The GMP footer is printed after several measurement results by pressing and
holding the p key, e.g. for component printing (menu item “For multiple
application results“). In this case, the T symbol remains visible until the GMP footer
is printed.
If you toggle to a dierent platform while a GMP printout of several measured
results is being generated, the GMP footer for the platform used up to that point is
generated when you press the n key. The GMP header for the other platform is
included on the next printout generated.
A GMP-compliant printout is generated automatically at the conclusion of
calibration/adjustment and linearization routines, as well as when you set or delete
a preload.
If you use a label printer for GMP-compliant printouts, you may nd that a single
label is not long enough for the data printed. If this is the case, you can activate the
automatic form feed after each printout of a GMP header and measurement results.
The following provides sample GMP headers and footers (see “Sample Printouts“).
108 Operating Instructions Combics Complete Scales
Data Interfaces