t Use the “O“ soft key to open the CAL key function“ menu.
y The “CAL key function“ submenu is displayed.
t Use the “Q“ or “q“ soft key to select the corresponding menu item and conrm
with “l.“
y The menu item is marked by a circle “o.“
Note
The functions that can be congured in the “CAL key function“
submenu depends on the selected weighing platform and its conguration data.
Functions that cannot be accessed are not displayed in the selection screen.
Important Note!
The function set in the “CAL key function“ menu is carried out in the
normal weighing mode because when you exit the Setup menu the Service mode is
deactivated. To perform the function on a digital weighing platform (such as an
IS platform), however, it must be carried out in Service mode.
t The procedure is as follows: after selecting the desired function and exiting
the Setup menu, reactivate Service mode again and then exit the Setup menu
immediately by pressing the M key or the “
oo“ soft key.
y The scale is now in Service mode without this being displayed.
t Trigger the previously set function using the J key.
y The display shows “S_CAl:“ indicating that the scale is in Service mode.
t The corresponding menu item can be selected using the “Select“ soft key
and carried out using the J key.
t If you cancel the function using the ( key or by restarting the scale using
the e key, you will exit the Service mode.
External Calibration/Adjustment
with Factory-Dened Weight (Default Weight)
Conguration If not already selected (factory setting, marked by “o“ when active), select the
“CAL key function:Ext. cal./adj.; factory-
def. wt.
“ menu item (external adjustment with a factory-dened, standard
weight).
t Use the “Q“ or “q“ soft key to select this menu item and conrm with “l.“
y The menu item is marked by a circle “o.“
Note:
The menu items “Ext. cal./adj.; factory-def. wt.“
(external adjustment with a factory-dened, standard weight), “Ext. cal./
adj.; user-defined wt.
“ (external adjustment with
a user-dened weight) and “Key blocked“ can also be accessed without
activating the Service mode.
Operating Instructions Combics Complete Scales 39
Configuring Weighing Platforms