t Press J to apply linearization weight 1 or press ( to cancel the
linearization function.
y After the linearization weight 1 has been saved you will be prompted to place
the second linearization weight on the weighing pan.
t Repeat the procedure for all required linearization weights.
y After the last linearization weight has been saved you will be prompted to
remove any load from the weighing pan.
t Remove all linearization weights from the weighing platform.
y The zero point is applied automatically after a brief time. The indicator
automatically switches to weighing operation.
When linearization is complete, the GMP-compliant printout shown on the left
is generated.
t Slide the menu access switch to the right (= “closed“ position).
--------------------
24.02.2016 10:15
Model CAW3E
Ser.no. 12345678
Vers. H0 111.031115
BVers. 01-63-02
Ser.no. C 12345678
--------------------
Linearization
Wt.1 + 1.500 kg
Wt.2 + 3.000 kg
Wt.3 + 4.000 kg
Wt.4 + 6.000 kg
completed
--------------------
24.02.2016 10:17
Name:
--------------------
Operating Instructions Combics Complete Scales 49
Configuring Weighing Platforms

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea Intec CAS-SPC Series and is the answer not in the manual?

Minebea Intec CAS-SPC Series Specifications

General IconGeneral
BrandMinebea Intec
ModelCAS-SPC Series
CategoryScales
LanguageEnglish

Related product manuals