29
Item Function No.
Setting
value
Contents
0 1 scale interval
1 2 scale interval
●2 5 scale interval
3 10 scale interval
4 20 scale interval
Setting of scale interval of
the second rang
CF-41
5 50 scale interval
Boundary value setting in the
second range
CF-42
000000
~999999
●003000
0 1 scale interval
1 2 scale interval
2 5 scale interval
●3 10 scale interval
4 20 scale interval
Setting of scale interval of
the third rang
CF-43
5 50 scale interval
Boundary value setting in the
third range
CF-44
000000
~999999
●006000
●0 Display interlock
1 Gross value Analog output target CF-70
2 Net value
Display value at the analog
output of DC4 mA(DC0 V)
CF-71
0~999999
●0
Unit : 1D
Display value at the analog
output of DC20 mA(DC10 V)
CF-72
0~999999
●10 000
Unit : 1D
Calibration data of the scale
interval (Reference use)
CF-90 ●00001
Calibration data of the
maximum capacity(Reference
use)
CF-91 ●10000
Calibration data of the weight
(Reference use)
CF-92 ●10000
Calibration data of zero
mV/V (Reference use)
CF-93 ●0.0000
Calibration data of span
mV/V (Reference use)
CF-94 ●0.5000
Stability detection time width
at calibration
CF-97
00~99
●05
Unit : 0.1 s
00 : OFF of stability detection at
calibration
Digital linearization clear CF-98
The data of the compensation by the
digital linearization is cleared.
Memory clear CF-99
C function set content is returned to the
default setting
● has been set as default.