45
8-7. Automatic printing
Automatic printing is used when the print output is automatically done whenever the load hangs to the
measuring section by using the two-wires method serial interface, the BCD output, and the RS-232C interface.
The load display enters within the range of the output effective, and data is output once at the stable.
8-7-1. Operating condition of automatic printing
The operating condition of automatic printing can be selected by function F-16.
The operating condition of automatic printing can be selectable from OFF, Data exceeding +5 counts and
Data out of the range from 5 counts to + 5 counts.
Default has selected OFF, and printing is made by the input of the key or external control input.
8-8. Accumulation
The accumulation function is a function to add the load data accumulating. Accumulating frequency and total
value are memorized.
8-8-1. Operating condition of accumulation function
The operating condition of accumulation function can be selected by function F-50.
The operating condition of accumulation can be selectable from OFF, Only the manual accumulation + and
“Both of the manual accumulation +/-”. “Only the automatic accumulation +” and “both of the automatic
accumulation +/-”.
Default has selected Both of the manual accumulation +/-.
8-8-2. Setting of accumulation prohibited zone
The accumulation prohibited zone is the zone which cannot accumulate, and the next accumulation can be
executed after the load value enters into Accumulation prohibited zone once.
The accumulation prohibited zone can be set by function F-51.
Setting range : 0099 Unit : D 00Accumulation always when stability is detected.
Default has set 05.
8-8-3. Execution of accumulation
Manual accumulation
Only + : When it is the stable condition and the key is pressed in the condition of accumulation
prohibited zone < load value, the accumulation is executed.
Both + and - : When it is the stable condition and the key is pressed in the condition of
accumulation prohibited zone < |load value|, the accumulating calculation is executed.
Automatic accumulation
Only + : When it detects the stable condition in the condition of accumulation prohibited zone < load
value, the accumulation is executed automatically.
Both + and - : When it detects the stable condition in the condition of accumulation prohibited zone
< |load value|, the accumulating calculation is executed automatically.
The next accumulation becomes in effective after the load value enters into Accumulation prohibited zone once.
When the setting of accumulation prohibited zone F-51 is 00, it becomes stability release once from the stable state, it accumulates the
load value when becoming an stable state again.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea CSD-904-EX and is the answer not in the manual?

Minebea CSD-904-EX Specifications

General IconGeneral
BrandMinebea
ModelCSD-904-EX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals