24
4-6. Calibration only of zero point
It is a function for the re-calibration only of the zero point when the amounts of the tare other than the
maximum capacity of the measuring section are changed.
4-6-1. Changeover of the calibration mode by zero point
Press the key to change from the measurement
mode to the stand-by condition.
is displayed when the key is
pressed with the key.
is displayed by pressing the key
after pressing the key twice.
is displayed after pressing the key
at three times, and entered into the calibration mode
only of the zero point.
Measurement mode
Stand-by condition
Calibration mode only of zero point