6-6. Power on zero
The power on zero function makes the display to a zero if the changes of display is within ±10 % of the
maximum capacity when the power is turn on, or in the stable condition in the operation of display turning on.
6-6-1. The operation of power on zero
Effective and Invalidity of the power on zero operation are selected by C function CF-15.
6-7. Tare weight cancellation
By pressing the key, the display of TARE and NET in the condition display section blinks and change
to zero.
To execute the tare weight cancellation clear, press the key when the gross value is zero. The tare weight
cancellation is cleared and GROSS of the condition display section lights on at the same time as becoming a
gross value display.
6-7-1. Operating condition of tare weight cancellation
The operating condition of tare weight cancellation is selected by C function CF-16.
Operation when it is steady or Operate unconditionally can be selected.
Default has selected Operation when it is steady.
The zero compensation is not accepted in the following condition when the setting of C function CF-15 is Effective power on zero and
when power on zero is starting.
When the display is not an stable condition, it becomes as all lighting.
When the measurement value exceeds ±10 % of the maximum capacity, is displayed. At this time, if the key
is pressed, it compulsorily becomes a load value display.
When the power on zero is executed, each data of tare weight cancellation , fixed tare weight cancellation , zero set and zero tracking
shall be all cleared.
The tare weight cancellation is not accepted when gross value is below the zero.
The tare weight cancellation is not accepted when the gross value is larger than the first range.
When the tare weight cancellation is executed with a fixed tare weight cancellation executed, a fixed tare weight cancellation is canceled,
and the tare weight cancellation is executed. (A fixed tare weight cancellation cannot be used together with the tare weight cancellation)
When the zero set is executed in the condition of gross value display, the tare weight cancellation will be cancelled.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea CSD-904-EX and is the answer not in the manual?

Minebea CSD-904-EX Specifications

General IconGeneral
BrandMinebea
ModelCSD-904-EX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals