43
8-4. Key lock function
This function sets whether the operation of each key in the front panel is effective or invalid to prevent the
various keys from the wrong operation.
8-4-1. Key lock
The function of F-08 sets whether the key operation of set,Accum./ ,PRESET TARE/ ,
TARE/ and NET/GROSS/ is effective or invalid.
The key lock function becomes effective only in the measurement mode.
Default is set Invalid.
8-5. Stability detection
When the change width of the load display is within the data width set by function F-10, and when the condition
continues for the time set by function F-11 or more, the stability detection is the function to judge the
measuring value being stable.
8-5-1. Data width of stability detection
The data width of stability detection can be set by function F-10.
Setting range : 09 Unit : 0.5D 0OFF
Default has set 4.
The data width is obtained in the display conversion by the below formula
Data width of stability detection Setting value of F-10 x 0.5 x The scale interval
When the setting of function F-10 is 4, and when the scale interval is D = 2, it will become as follows;
Data width stability detection = 4 x 0.5 x 2 = 4D

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea CSD-904-EX and is the answer not in the manual?

Minebea CSD-904-EX Specifications

General IconGeneral
BrandMinebea
ModelCSD-904-EX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals