36
ex.2 In case of net value cancelled 300 g of tare weight set in ex.1
The measurement from -300 to 300 g becomes the first range(scale interval 0.1 g).
The measurement from 300 to 600 g becomes the second range(scale interval 0.5 g).
The measurement from 600 to 700 g becomes the second range(scale interval 1 g).
6-10. Setting of the stability detection time width in calibration
The stability detection time width in calibration can be set by C function CF-97.
Setting range : 0099 Unit : 0.1 s
Default has set 05.
6-11. Digital linearizer clear
The compensation data set in the digital linearization can be cleared by the C function CF-98.
If the key is pressed with display, becomes a blinking display.
Press the key to discontinue the digital linearization at this point.
It becomes a stand-by condition, and the digital linearization is not clear.
is displayed If the key is pressed when displays blinking, and the
operation of digital linearization is completed.
6-12. Memory clear
A clear memory is executed by C function CF-99, and the setting contents of C function is returned to the
default setting.
If the key is pressed when displays blinking, becomes a blinking display.
Please press the key to discontinue the clear memory at this point.
It becomes a stand-by mode, and digital linearization is not clear.
If the key is pressed when displays blinking, the display section is turned off and
becomes to display, then the operation of memory clear is completed.
When it is used only within the first range (C function CF-40 : OFF), it is not necessary to set the range.
Please make the relations of the scale interval of the each range as the first rangethe secondd rangethe third range.
Please make the relations of the each boundary value of each range as the second range≦the third range≦the maximum capacity.
The range upper bound within the range of second is a maximum capacity for the switch by two ranges.
The range upper bound within the range of third is a maximum capacity for the switch by three ranges.
The comparator setting and the fixed tare weight cancellation setting become the scale interval set in the first range.
0g
300 g
600 g 1 000 g
目量 0.1 g
目量 0.5 g
目量 1 g
計量値
0g
-300 g
300 g
600 g 700 g
総量
正味量
Scale interval 0.1 g
Scale interval 0.5 g
Scale interval 1 g
Net value
Measurement value
Gross Value
Tare weight

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea CSD-904-EX and is the answer not in the manual?

Minebea CSD-904-EX Specifications

General IconGeneral
BrandMinebea
ModelCSD-904-EX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals