44
8-5-2. Time width of stability detection
The time width of stability detection can be set by function F-11.
Setting range : 0~9 Unit : 0.5 s 0:OFF
Default has set “2”.
8-6. Input of the fixed tare weight cancellation
The fixed tare weight cancellation function executes the tare weight cancellation according to the set value by
which digital is input.
8-6-1. Operating condition of the fixed tare weight cancellation
The operating condition of the fixed tare weight cancellation can be selected by the selection of funciton
F-15. Default has set “effective”.
8-6-2. Input of the fixed tare weight cancellation
If the key is pressed from the normal measurement
mode, it becomes the input display of the setting for
fixed tare weight cancellation.
As the value being memorized now is displayed while the
tare weight cancellation is setting, please set the value of
fixed tare weight cancellation by using the following keys.
: Change the value of changing digit.
: Select the changed digit.
: Change the display value into “0”.
: The displayed value is not memorized, and
it returns to the measurement mode.
: Memorize the display value and proceed to
the next step.
: The display is turned off without changing all
setting, and it becomes the stand-by
condition.
After changing, the key is pressed. A fixed tare
weight cancellation is executed, the load display becomes
“Net value = Gross value – fixed tare weight setting
value”, and “PRESET” and “NET” of the status display
lights on.
● When two range or three range switches have been selected by the automatic range switches, the data width of the stability detection
execute
the display conversion of the data width by the value of the first range scale interval (scale interval in the calibration).
● When two range or three range switches have been set by automatic range switches, the setting range where a fixed tare weight
cancellation can be input is within the first range.
● When a fixed tare weight cancellation is executed with the tare weight cancellation executed, the tare weight cancellation is cleared, and a
fixed tare weight cancellation is executed. (The tare weight cancellation cannot be used together with a fixed tare weight cancellation.)
● When the zero set is executed in the condition of gross value display, the fixed tare weight cancellation will be cleared.
Measurement mode
Measurement mode