13-3-12. Data format of command mode
Reading out the load data (Host CSD-904-EX)
When command number is from 20 to 25
2
4 6
@ 0 0 2 2 Terminator
Return (CSD-904-EX Host)
2
4
6
8
10
12
14
@ 0 0 2 2
Terminator
2
4
6
8
10
12
14
@ 0 0 2 2
1 2 3 Terminator
2
4
6
8
10
12
14
@ 0 0 2 2
0 . 1 2 3 Terminator
2 4 6 8 10 12 14
@ 0 0 2 2
OL Terminator
2 4 6 8 10 12 14
@ 0 0 2 2
OL Terminator
The address becomes 00 fixation for the RS-232C interface.
The load data enters from the right.
The sign of - for minus and + for plus is entered.
The load data performs zero suppress.
The decimal point adheres to a specified position when the decimal point is set with F-46 and when decimal point is specified with CF-01.
OL is appeared in the load data section at the over load.
The empty sections are all spaces.
26
Address
0031
Command No.
Command No.
20
21
22
23
24
25
Operation
Reading out load data
Reading out gross value
Reading out net value
Reading out tare value
Reading out accumulation total value
Reading out accumulation count
Reading out load display data with
state
Address
0031
Command No.
Sign of load data
Load data
Address
0031
Address
0031
Address
0031
Address
0031

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea CSD-904-EX and is the answer not in the manual?

Minebea CSD-904-EX Specifications

General IconGeneral
BrandMinebea
ModelCSD-904-EX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals