21
Installation
Chillgard 5000
US
4.4.5 Exhaust Venting
The use of exhaust tubing reduces pump noise from the device.
Make sure to vent the exhaust of the device to the outside atmosphere (preferred) or a safe area.
Do not use inline or end-of-line filters in the exhaust line.
If possible, route the exhaust line so venting occurs to the outside atmosphere. Make sure to protect
the end of the exhaust tube from elements such as water, dirt, snow, ice, and insects, which can cause
clogs and prevent efficient venting.
If exhaust venting occurs indoors, make sure to route the exhaust line to:
• An area that is not monitored for refrigerant gases
• An area away from personnel
WARNING!
Make sure to remove the cap from the exhaust port before operating the device. Failure to do so will
create back pressure that can result in inaccurate readings and damage to internal components.
Failure to follow these warnings can result in serious bodily injury or death.
4.5 Wiring and Grounding
The Chillgard Wiring Diagram (Fig. 11) gives the details of the wiring requirements for the device.
WARNING!
Make sure the device chassis is connected to the earth ground at the ground bar.
Failure to follow these warnings can result in serious bodily injury or death.
NOTICE
Installations that require conformity to the CE mark must have a connection between the
Chillgard 5000 device and a nearby earth ground potential.
To achieve this connection, install a 10 AWG or larger copper wire between the grounding lug terminal
strip on the right side of the Chillgard 5000 device and the grounding point (protective earth). The wire
length cannot be longer than 6 ft (2 m).