50
Routine Operation
Chillgard 5000
US
7.7.3 Faults
The last 100 events show in chronological order. When there are 100 events in the list and a new
event occurs, the oldest event is deleted to make space for the new event to show.
Use the scroll bar to move through the list of events.
7.7.4 Calibration
The last 20 calibrations show in chronological order. When there are 20 calibrations in the list and a
new calibration occurs, the oldest calibration is deleted to make space for the new calibration to show.
Use the scroll bar to move through the list of events.
The "Faults" tab shows a list of all fault events for the
device.
Select any event on the "Faults" tab to see a pop-up with
event details. The detailed view shows the board location,
if applicable, and the software error code.
The "Calibration" tab shows a list of calibration attempts for
the device.
Select any calibration event to see a pop-up with details.
NOTE: The "As Found" value represents the performance
reading before calibration is performed. The "As Left" value
represents the performance reading after calibration is
performed.