38
Calibration
Chillgard 5000
US
6.3 Zero Gas Calibration
(1) Connect the zero-gas scrubber tubing to the calibra-
tion port.
(2) On the device touchscreen, select "Next".
A progress bar shows during the zero gas calibration.
(3) When zero gas calibration is complete, the "Results"
screen shows. Select "Continue" to go to the next
step.
(4) On the "Calibration Complete" screen, do one of the
following:
a) To continue with span gas calibration, select "Span
Gas Calibration". Go to Step (1) in the Span Gas
Calibration section below.
b) To complete zero gas calibration without continuing
to the span gas calibration, select "Done", and go to
the next step.
c) To return to the "Results" screen, select "Back".
(5) On the bottom of the device, install the pin in the cali-
bration port.
(6) Remove the tubing from the calibration port.