Manual Inflation
1 Set the e360 ventilator to Volume Control SPONT.
2 Press the “Manual Inflation button.
3 Verify that a manual inflation is delivered while the button is
pressed.
Inspiratory Tidal Volume
1 Set the e360 ventilator to Volume Control A/CMV.
2 Set Flow to 30 L/min.
3 Press “Setup & Calibration button on the Control Panel, and
then press “
Patient Setup button on the LCD touch screen.
4 Press “Circuit Type button on the LCD touch screen, and set
Circuit T
ype
to HME by
Touch-Turn-Accept method.
5 Press “Main Screen button on the Control Panel, and then
press “
Numeric button on the LCD touch screen.
6 Set the calibration analyzer to volume measurement in BTPS
mode, and connect it in the inspiratory limb of the patient
circuit.
7 Set Tidal Volume to 250 mL. Verify the value of VTI displayed
on
Numeric
screen, the tidal volume reading measured
by the calibration analyzer and the
Tidal Volume setting are
within tolerance of ± 10% among each other. Record readings
on the
T
est Record.
8 Repeat step 7 with Tidal Volume settings of 500 mL, and 1L.
Expiratory Tidal Volume
1 Connect the calibration analyzer in the expiratory limb of the
patient circuit.
SECTION 6
SER360 A1106 6-11