EasyManua.ls Logo

Newport e360 - Regulator and Solenoid Assembly BKT2105 A

Newport e360
129 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Removal and Replacement of Regulator and Solenoid Assembly BKT2105A
Figure 4-15 Regulator and Solenoid Assembly BKT2105A
Refer to figure 4-15 and the System Pneumatic Diagram in
Appendix C, and follow these steps:
1 Perform steps above for removal of Pneumatics Panel
Assembly SVO2101A.
2 Remove the yellow tubing from the
Air Inlet Block BLK2115M,
and remove the blue tubing from the Oxygen Inlet Block
BLK2115M. Refer to Fig 4-14.
3 Remove the clear tubing from the top port of the pressure
transducer XD105 (machine 1 pressure) on the analog board
PCB2104A, and remove the clear tubing from the exhaust port
of the rezero solenoid SOL1501P for machine pressure 1.
4 Remove the green tubing from the bottom port of the pressure
transducer XD106 (machine 2 pressure) on the analog board
PCB2104A, and remove the green tubing from the exhaust
port of the rezero solenoid SOL1501P for machine pressure 2.
5 Remove the blue tubing, one end of which is connected to the
Exhalation Servo Valve VLV1806P, from the other end, which
is connected to a tee connector
.
6 Remove the green tubing, one end of which is connected to
the IN port of the Emergency Relief Solenoid SOL1501P
, from
the other end, which is connected to a cross tee connector.
COMPONENT REMOVAL & REPLACEMENT
2
screws, 1 each side securing
BKT2105A to e360
R
egulator and solenoid
assembly BKT2105A
C
rossover solenoid
SOL1501P
E
xhalation Servo
Vavle VLV1806P
Emergency Relief
Solenoid SOL1501P
Machine 1 Pressure
Rezero Solenoid
SOL1501P
Machine 2 Pressure
Rezero Solenoid
SOL1501P
Safety Regulator
REG1802P
Exhalation Servo
Regulator REG1701P
4-16 SER360 A1106

Table of Contents

Other manuals for Newport e360

Related product manuals