Exhalation Flow Sensor: d9
1 Turn OFF the e360 ventilator.
2 Turn ON the e360 ventilator and wait until the “Ventilation
Standby
screen is displayed.
3 Connect the patient circuit to the e360 ventilator as shown in
Figure 6-1.
4 Occlude the patient circuit at the Y-piece using a cap. A filter
and test lung is not required for this calibration.
5 Press “CIRCUIT CHECK on LCD touch screen to perform
the circuit check. Confirm that the circuit check passes.
6 Press “SENSORS on LCD touch screen, and press “Flow
Sensor
to start Exhalation Flow Sensor calibration. It may
require performing several times in order to pass the
Exhalation Flow Sensor calibration.
7 Turn OFF the e360 ventilator and turn ON while pressing and
holding “
Accept button to enter Diagnostics Mode.
8 Set the e360 ventilator to Diagnostics Mode “d9”.
9 With patient circuit occluded, press “Manual Inflation button
to start Exhalation Flow Sensor calibration in Diagnostics Mode
.
The Message ID will change to “
63”.
10 Wait for several minutes to complete the calibration until the
Message ID changes to “41”.
Exhalation Servo Valve: d4
1 Set the e360 ventilator to Diagnostics Mode “d4”.
2 With patient circuit occluded, press the “Manual Inflation
button once. The “Message ID” will change to “
35”.
3 Press the “Manual Inflation” button again to start the
calibration, now the “
Message ID will change to “15”.
4 Wait for several minutes to complete the calibration until the
“Message ID” changes to “
20”.
5-14 SER360 A1106
CALIBRATION PROCEDURES