Exhalation Flow Sensor: d9
1 Turn OFF the e360 ventilator.
2 Turn ON the e360 ventilator and wait until the “Ventilation
Standby
” screen is displayed.
3 Connect the patient circuit to the e360 ventilator as shown in
Figure 6-1.
4 Occlude the patient circuit at the Y-piece using a cap. A filter
and test lung is not required for this calibration.
5 Press “CIRCUIT CHECK” on LCD touch screen to perform
the circuit check. Confirm that the circuit check passes.
6 Press “SENSORS” on LCD touch screen, and press “Flow
Sensor
” to start Exhalation Flow Sensor calibration. It may
require performing several times in order to pass the
Exhalation Flow Sensor calibration.
7 Turn OFF the e360 ventilator and turn ON while pressing and
holding “
Accept” button to enter Diagnostics Mode.
8 Set the e360 ventilator to Diagnostics Mode “d9”.
9 With patient circuit occluded, press “Manual Inflation” button
to start Exhalation Flow Sensor calibration in Diagnostics Mode
.
The Message ID will change to “
63”.
10 Wait for several minutes to complete the calibration until the
“
Message ID” changes to “41”.
Exhalation Servo Valve: d4
1 Set the e360 ventilator to Diagnostics Mode “d4”.
2 With patient circuit occluded, press the “Manual Inflation”
button once. The “Message ID” will change to “
35”.
3 Press the “Manual Inflation” button again to start the
calibration, now the “
Message ID” will change to “15”.
4 Wait for several minutes to complete the calibration until the
“Message ID” changes to “
20”.
5-14 SER360 A1106
CALIBRATION PROCEDURES