Removal and Replacement of Main Flow Outlet Block Assembly BLK2108A
Figure 4-17 Door DOR2112M
Figure 4-18 Outlet Block BLK2108A
Refer to figure 4-17 and 4-18 and follow these steps:
1 Remove the 2 screws securing DOR21
12M to the front of the
e360 ventilator
.
COMPONENT REMOVAL & REPLACEMENT
Door DOR2112M
Remove 2 screws
Remove 2 screws
Oxygen sensor cable CBL2109A Outlet block BLK2108A
4-18 SER360 A1106