Emergency Relief Diaphragm
To replace the Emergency Relief Diaphragm, follow the
procedures given in the Preventive Maintenance at the beginning
of this section.
Emergency Intake Diaphragm
Figure 2-8 Emergency Intake Diaphragm Replacement
1
Remove Top Cover following procedure in Section 4 of this
manual, “Removal of
Top cover”.
2 Remove Pneumatics Panel following procedure in Section 4 of
this Manual, “Removal of Pneumatics Panel Assembly”.
3 Remove Inspiratory Flow Sensors following procedure in
Section 4 of this manual, “Removal of Inspiratory Flow
Sensors”.
4 Remove two Emergency Intake Valve retaining screws, see
Figure 2-8
above.
5 Remove the Emergency Intake V
alve using Needle Nose
Pliers.
6 Replace the Emergency Intake Diaphragm (VLV100P) and
reinstall assembly by reversing procedure above.
SECTION 2
Mixing Block
Retaining Screws (2)
Mixing Block
Emergency
Intake Valve
Emergency Intake
Valve Retaining
Screws (2)
SER360 A1106 2-9