Removal and Replacement of Inhalation Outlet Block and Mixing Block Assembly
Figure 4-20 Inhalation Outlet block and mixing block
Refer to figure 4-20 and follow these steps:
1 Perform above steps for removal of Oxygen Sensor R125
P03-002.
2 Remove 2 screws on top of mixing block BLK2107M.
3 Remove both the mixing block and inhalation block from the
e360 ventilator.
4 Remove 2 screws and washers securing mixing block
BLK2107M to inhalation outlet block BLK2117M and separate
the two.
5 To reinstall, reverse above steps 2 through 5.
COMPONENT REMOVAL & REPLACEMENT
Inhalation outlet block
“A” BLK2117M
Mixing block
BLK2107M
R
emove 4 screws
and washers
4-20 SER360 A1106