WARNING
To maintain grounding integrity, connect only to a hospital
grade receptacle. Always disconnect power supply before
servicing the e360 ventilator.
DANGER: there is a risk of explosion if used in the presence of
flammable anesthetics.
Before returning to patient use, the e360 ventilator must pass
the operational verification procedure.
All e360 ventilator service or repair must be performed by a
technician authorized and trained by Newport Medical
Instruments.
Use extreme caution when working inside the e360 ventilator
while it is connected to a power source.
GENERAL CAUTIONS
Caution
Use standard antistatic techniques when working inside the
e360 ventilator or handling any electronic parts.
Clean all external parts of the e360 ventilator prior to service.
Use only dry, clean compressed air and medical grade oxygen.
Water in the air or oxygen supply can cause equipment
malfunction and damage.
Mains voltage must correspond to the voltage range selected
on the power module of the e360 ventilator. Always replace an
open fuse with one of correct type and rating.
Do not place containers of liquids near the e360 ventilator.
Liquids that get into the e360 ventilator can cause equipment
malfunction or damage.
NOTE: Use the tools specified in the manual to perform specific
Procedures.
WARRANTY
The e360 ventilator comes with a two (2) year conditional
warranty
.
The warranty covers any defect or malfunction that
occurs due to normal use. The warranty does not cover any
INTRODUCTION
1-2 SER360 A1106