7 Remove 2 screws and washers securing BKT2105A to the
e360 ventilator.
8 To reinstall, reverse the above steps.
Removal and Replacement of FTD2100P and FTD2101P Inspiratory Flow Sensors
Figure 4-16 Inspiratory Flow Sensors
Refer to figure 4-16 and follows these steps:
1 Perform steps above for removal of pneumatics panel
assembly SVO2101A.
2 Remove cable CBL2116A from connector J1 on the flow
sensor.
3 Remove the flow sensors from mixing block BLK2107M by
pulling the flow sensor straight out of the mixing block.
4 To reinstall, reverse steps 2 and 3.
Note: If either of the flow sensors are replaced, calibrations D1
and /or D2 will need to be performed.
SECTION 4
Oxygen flow
sensor
FTD2101P
M
ixing block
BLK2107M
C
able CBL2116A
Connected to J1 on
flow sensors
A
ir flow sensor
FTD2100P
SER360 A1106 4-17