Figure 4-24 Removal of MNF2110A
Removal and Replacement of Exhalation Flow Sensor Board PCB2103P
Refer to figure 4-25 and follow these steps.
Perform above steps for removal of Exhalation Manifold
MNF2110A.
1 Remove cable CBL2122A from K1 and CBL2124P from K5 on
PCB2103P.
2 Remove 2 screws and washers securing PCB2103P to the
Exhalation Manifold MNF2110A.
3 Remove PCB2103P.
4 To reinstall, reverse above steps 1 through 4.
Note: There is a small hole adjacent to K1 on PCB2103P, which is
to polarize the connector
.
COMPONENT REMOVAL & REPLACEMENT
Exhalation valve drive line Machine 2 pressure line
4-24 SER360 A1106