OPERATIONAL VERIFICATION PROCEDURE
The Operational Verification Procedure ensures that the Newport
e360 Ventilator is in proper operating condition.
WARNING Do not use the e360 ventilator unless it passes the
Operational Verification Procedure.
Test Equipment
Electrical safety test (EST) analyzer
Certified calibration analyzer manufactured by TSI, Inc. or
equivalent
Oxygen Analyzer
Newport reusable patient breathing circuit (PBC340A) or
equivalent
Adult test lung (LNG800P), or adult test lung with RP20 airway
resistor (LNG600A) or equivalent
Adjustable (0-60 psig) dry clean regulated medical grade air
and oxygen gas supplies
A cap to occlude the patient circuit Y-piece (CAP100P)
or equivalent
Caution The accuracy of all test equipment (electronic or
pneumatic) used for verification or calibration procedures
should be certified annually by a testing laboratory before use.
SECTION 6
SER360 A1106 6-1