3 Remove and replace the inlet filter and O-ring.
4 Reinstall the filter holder and reassemble the water trap
assembly.
Figure 2-2 Exhalation Valve Adapter Replacement
1
Access the Exhalation Valve
Adapter by lifting the Retaining
Latch and removing the Exhalation Valve and Exhalation Flow
Sensor, see Figure 2-2 above.
2 To remove the silicon Exhalation Valve Adapter, use your finger
to grasp the Adapter and pull it straight out.
3 To replace the Exhalation Valve Adapter, orient the Adapter lip
towards the inside of the instrument and press into place.
4 Ensure that the Adapter is properly seated before re-installing
the Exhalation Valve and Exhalation Flow Sensor.
MAINTENANCE, OVERHAUL & SOFTWARE UPGRADE
Valve
Adapter
Adapter Lip
Retaining
Latch
ADP2105M
Exhalation Valve Adapter
2-4 SER360 A1106