Removal of Key Components
Unless instructed otherwise, reinstall parts by reversing the
disassembly procedure.
Preparation
Before performing any repair or service on the e360 ventilator do
the following:
1 Disconnect the power.
2 Shut off or disconnect the air and oxygen gas supplies.
3 Put on an ESD wrist strap and perform procedures in an ESD
safe environment.
Tools Required
• #0 Philips Screw Driver
• #1 Philips Screw Driver
• #2 Philips Screw Driver
• #1 Flat Screw Driver
• 5/64” Hex Key
• 7/32” Nut Driver or Wrench
• 11 mm Nut Driver or Wrench
• Torque Wrench set for 8 inch/pounds
• Anti Static Wrist Strap
• Anti Static Parts Bin
WARNING To avoid possibility of electric shock, do not touch
the power supply transformer when servicing the ventilator.
These instructions are intended for use only by a Newport
Medical Instruments factory-trained technician. Do not perform
any unauthorized modifications or repairs to the ventilator or its
components.
Caution To avoid damaging equipment, always use standard
electrostatic discharge (ESD) precautions, including an ESD
wrist strap, when servicing the ventilator.
Note: If the Main Board, Display Board or Single Board computer
components are replaced the e360 software must be reloaded.
Follow the Software Upgrade Procedure in Section 2 to load
software. Contact Newport
T
echnical Service if you need a
software kit.