Current Leakage test
1 Power the e360 ventilator ON.
2 Perform the current leakage check and record the result on
the OVP record sheet. To pass, the measured current leakage
must be ? 100 µA.
OPERATIONAL CHECKS
Ensure the e360 ventilator Control Panel is adjusted to the
settings shown in Table 6-1.
NOTE: Adjustments are made with the Touch-Turn-Accept
method.
Table 6-1. Standard Test Setting
SECTION 6
Control Setting
Patient Type Adult
Power ON
Breath Type Volume Control
Mode A/CMV
F
I
O
2
.60
Flow 30 L/min
Tidal Volume 500 mL
Resp Rate 10 b/min
Pressure Limit 5
Pressure Support 0
PEEP 0
Trig Pressure 3.0 cmH
2
O
MVE Alarm
Low at 1 L/min, High at 50 L/min
Pause 0 sec
Paw Alarm Low at 10, High at 75
Leak Comp. OFF
Circuit Type HVE
NOTE: Pressure readings should be set to either cmH
2
O or mbar
SER360 A1106 6-3