3 Grab the handles on the sides of the e360 ventilator, pull the sides
away from the chassis and slide the cover off of the e360 ventilator.
4 To reinstall, reverse above steps 2 and 3.
Removal and Replacement of Front Panel Assembly DSP2102A
Figure 4-3 Front Panel Assembly DSP2102A
Refer to 4-3 and follow these steps:
1 Perform steps 1 and 2 for removal of top cover.
2 Locate the four screws securing the front panel assembly to
the chassis of the e360 ventilator and remove.
3 Carefully pull forward and away the front panel assembly from
the e360 ventilator chassis. Locate and remove the following cables.
• CBL2101A from J2 on PCB2100A
•
CBL21
18A
from J13 on PCB2100A
•
CBL21
10A
from J8 on PCB2100A
• CBL2113A from J10 on PCB2100A
•
Air and oxygen servo valve cables from J705 and J706 on
PCB2100A
• CBL2107A from J708 on PCB2100A
• CBL2122 from J9 on PCB2100A
• CBL2105A from connection LCD1 on SBC2100P
• 3 wire cable of CBL2112A from connection TS1 of
SBC2100P
.
TS1 is a 6 pin connector located adjacent to
the speaker SPK1.
COMPONENT REMOVAL & REPLACEMENT
Remove 4
s
crews, 2
each side
4-4 SER360 A1106