EasyManua.ls Logo

Newport e360 - Oxygen Servo Valve

Newport e360
129 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Caution: Do not attempt to perform Air Servo Valve calibration
unless the air gas supply is regulated to 50 ± 2 psig and capable
of supporting a minimum of 180 L/min of constant flow.
1 Set the e360 ventilator to Diagnostics Mode “d1”.
2 Disconnect the patient circuit from the e360 ventilator.
3 Press the “Manual Inflation button to initiate Air Servo Valve
calibration. The e360 ventilator will go through a calibration
maneuver that slowly increases and then decreases flow twice.
4 The value of “TSI Air Flow displayed on LCD touch screen
will slowly change as the delivered flow changes.
5 Wait for several minutes to complete the calibration until the
Message ID changes to “6”.
Note: If the “Message ID shows “7” during this calibration, the Air
Servo Valve needs to be replaced.
Oxygen Servo Valve: d2
Caution: Do not attempt to perform Oxygen Servo Valve
calibration unless instructed by a NMI Representative or for one
of the following conditions:
Replacement or readjustment of Oxygen Inlet Regulator
Replacement of Oxygen Servo Valve
Replacement of Analog PCB
Replacement of Main PCB
Caution: Do not attempt to perform Oxygen Servo V
alve
calibration unless the oxygen gas supply is regulated to 50
±2
psig and capable of supporting a minimum of 180 L/min of
constant flow.
1 Set the e360 ventilator to Diagnostics Mode “d2”.
2 Disconnect the patient circuit from the e360 ventilator.
3 Press the “Manual Inflation button to initiate Oxygen Servo
Valve calibration. The e360 ventilator will go through a
calibration maneuver that slowly increases and then
decreases flow twice.
4 The value of “TSI O
2
Flow displayed on LCD touch screen
will slowly change as the delivered flow changes.
5 Wait for several minutes to complete the calibration until the
Message ID changes to “10”.
Note: If the “Message ID shows “7” during this calibration, the
Oxygen Servo Valve needs to be replaced.
SECTION 5
SER360 A1106 5-13

Table of Contents

Other manuals for Newport e360

Related product manuals