2 Set Tidal Volume to 250 mL. Verify that the value of VTE
displayed on
Numeric
screen is between 212 – 259 mL, and
verify that the tidal volume reading measured by the
calibration analyzer and the
Tidal Volume setting are within
tolerance of +
10%. Record readings on the Test Record.
3 Set Tidal Volume to 500 mL. Verify that the value of VTE
displayed on
Numeric
screen is between 423 – 517 mL, and
verify that the tidal volume reading measured by the
calibration analyzer and the
Tidal Volume setting are within
tolerance of +
10%. Record readings on the Test Record.
4 Set Tidal Volume to 1.00 L. Verify that the value of VTE
displayed on
Numeric
screen is between 840 mL – 1.03 L, and
verify that the tidal volume reading measured by the
calibration analyzer and the
Tidal Volume setting are within
tolerance of +
10%. Record readings on the Test Record.
Alarm Silence
1 Create an alarm condition by changing the settings and wait
for the alarm to sound.
2 Press the “Alarm Silence” button and verify the indicator
lights up, and the audible alarm stops.
3 Press the “Alarm Silence” button again and verify the
indicator turns of
f, and the audible alarm resumes.
4 Return e360 ventilator to standard test setting. Press Reset
button to clear alarm indicators.
5 Check off on the record sheet.
Loss of Power Alarm and Battery Operation
1 Disconnect the power cord from the AC power source.
2 Verify that the messages AC Power Loss and Battery Back Up
are displayed in the message window
. V
erify that the audible
alarm sounds, the Power Fail indicator flashes, and the
Internal Battery indicator lights.
3 V
erify that at least 3 bars of the charge bar graph are lit.
4 Reconnect the power cord to the
AC power source.
6-12 SER360 A1106
OPERATIONAL VERIFICATION PROCEDURE