Inhalation Outlet Check Valve
Figure 2-9 Inhalation Outlet Check Valve
1
Remove the Mixing Block by removing the two retaining
screws, see Figure 2-8.
2 Remove the Inhalation Adapter from the Inhalation Outlet
Assembly
, see Figure 2-5.
3 Insert finger and push out the Inhalation Outlet Check V
alve,
see Figure 2-9.
4 Replace the Inhalation Outlet Check Valve Diaphragm
(VL
V2100M).
5 To reinstall components and assembly, reverse above
procedure.
Internal Battery
T
o remove/install the Internal battery
, follow the procedures in
Section 4 of this manual, “Removal of Internal Battery”.
Caution: T
o avoid discharging battery voltage, do not allow
any metal object (including tools) to touch battery connectors.
Warning: Observe correct polarity when reconnecting battery
connectors.
Tubing
The tubing inside the e360 ventilator does not need to be replaced
at any predetermined time interval; however, Newport is aware
that tubing may occasionally need replacing. During the overhaul
procedure, carefully inspect all tubing for degradation, cracks, or
MAINTENANCE, OVERHAUL & SOFTWARE UPGRADE
Inhalation Outlet
Check Valve
2-10 SER360 A1106