Circuit Leak Test
Figure 6-3 Circuit Leak Test
NOTE:
Circuit Check is available in Ventilation Standby only.
1 Remove test lung if present and occlude the end of the
patient circuit Y-Piece using CAP100P.
2 Power ON the e360 ventilator or press “Setup & Calibration
button on the Control Panel.
3 Press the “Circuit Check button on the display and follow
the screen instructions.
4 After the test is completed, verify that the message display
window reads
PASSED. Check off the record sheet and
reattach the test lung.
5 To exit Ventilation Standby condition, press the “START
VENTILA
TING
button on LCD touch screen.
Loss of Gas Alarm
1 Lower the air gas supply pressure to 20 psig. Verify that there
is no Loss of Gas alarm.
2 Lower the air gas supply pressure to 8 psig. V
erify that the
audible alarm sounds, the amber LED on the top is flashing,
and the message display reads
Air Supply Loss.