Front Panel LED Check
Figure 6-2 Front Panel Controls & Indicators
1
Toggle the power switch located on the back of the e360
ventilator, to the OFF position.
2 Press down the “Accept button located next to the
Adjustment Knob.
3 While pressing the “Accept button, toggle the power switch
OFF then to the ON position.
This puts the e360 ventilator into
Diagnostics Mode. Release the ”
Accept button.
4 Verify that all the Control Panel display lights are lit except for
the Alarm Lamp on top of the ventilator.
6-4 SER360 A1106
OPERATIONAL VERIFICATION PROCEDURE
Manual Inflation
Encoder
Adjustment Knob