2 Remove 2 screws securing BLK2108A to the e360 ventilator.
3 Pull the BLK2108A away from the e360 ventilator and remove
2 tubes.
4 To reinstall, reverse above steps 2 and 3.
Removal and Replacement of Oxygen Sensor R125P03-002
Refer to figure 4-18 and 4-19 and follow these steps.
1 Remove the Oxygen Sensor cable CBL2109A from the
Oxygen Sensor R125P03-002 by unscrewing the connector
on the cable.
2 Remove the Oxygen Sensor by unscrewing the Oxygen
Sensor counterclockwise.
3 To reinstall, reverse above steps 1 and 2.
Note: If the Oxygen Sensor is installed too tightly to be removed,
place a Newport reusable circuit on the oxygen sensor to obtain
better friction while unscrewing the Oxygen Sensor.
Figure 4-19 Use circuit to loosen Oxygen Senor
SECTION 4
SER360 A1106 4-19